TRANG CHỦ GIỚI THIỆU
HỆ THỐNG QUẢN LÝ TUYỂN DỤNG LIÊN HỆ
 
Previous
DANH MỤC SẢN PHẨM
Liên Kết Web Site
Lượt Truy Cập
   SẢN PHẨM: 
SITRANS TH200



Mã sản phẩm: 7NG3211-1NN00
Tên sản phẩm: SITRANS TH200

Ultra flexible - with the universal SITRANS TH200 transmitter

 

Input

 

Resistance thermometer

 

Measured variable

Temperature

Sensor type

 
  • To IEC 60751

Pt25 … Pt1000

  • To JIS C 1604; a = 0.00392 K-1

Pt25 … Pt1000

  • To IEC 60751

Ni25 … Ni1000

  • Special type

over special characteristic (max. 30 points)

Sensor factor

0.25 ... 10 (adaptation of the basic type, e.g. Pt100 to version Pt25 ... Pt1000)

Units

°C or °F

Connection

 
  • Standard connection

1 resistance thermometer (RTD) in 2-wire, 3-wire or 4-wire system

  • Generation of average value

2 identical resistance thermometers in 2-wire system for generation of average temperature

  • Generation of difference

2 identical resistance thermometers (RTD) in 2-wire system (RTD 1 – RTD 2 or RTD 2 – RTD 1)

Interface

 
  • Two-wire system

Parameterizable line resistance ≤ 100 Ω (loop resistance)

  • Three-wire system

No balancing required

  • Four-wire system

No balancing required

Sensor current

≤ 0.45 mA

Response time

≤ 250 ms for 1 sensor with open-circuit monitoring

Open-circuit monitoring

can be switched off

Short-circuit monitoring

can be switched off (value is adjustable)

Measuring range

Parameterizable (see table "Digital measuring errors")

Min. measured span

10 °C (18 °F)

Characteristic curve

Temperature-linear or special characteristic

Resistance-based sensors

 

Measured variable

Actual resistance

Sensor type

Resistance-based, potentiometers

Units

Ω

Connection

 
  • Normal connection

1 resistance-based sensor (R) in 2-wire, 3-wire or 4-wire system

  • Generation of average value

2 resistance-based sensors in 2-wire system for generation of average value

  • Generation of difference

2 resistance thermometers in 2-wire system
(R1 – R2 or R2 – R1)

Interface

 
  • Two-wire system

Parameterizable line resistance ≤ 100 Ω (loop resistance)

  • Three-wire system

No balancing required

  • Four-wire system

No balancing required

Sensor current

≤ 0.45 mA

Response time

≤ 250 ms for 1 sensor with open-circuit monitoring

Open-circuit monitoring

can be switched off

Short-circuit monitoring

can be switched off (value is adjustable)

Measuring range

Parameterizable, max. 0 ... 2200 Ω (see Table "Digital measuring errors")

Min. measured span

5 Ω ... 25 Ω (see Table "Digital measuring errors")

Characteristic curve

Resistance-linear or special characteristic

Thermocouples

 

Measured variable

Temperature

Sensor type (thermocouples)

 
  • Type B

Pt30Rh-Pt6Rh to DIN IEC 584

  • Type C

W5 %-Re acc. to ASTM 988

  • Type D

W3 %-Re acc. to ASTM 988

  • Type E

NiCr-CuNi to DIN IEC 584

  • Type J

Fe-CuNi to DIN IEC 584

  • Type K

NiCr-Ni to DIN IEC 584

  • Type L

Fe-CuNi to DIN 43710

  • Type N

NiCrSi-NiSi to DIN IEC 584

  • Type R

Pt13Rh-Pt to DIN IEC 584

  • Type S

Pt10Rh-Pt to DIN IEC 584

  • Type T

Cu-CuNi to DIN IEC 584

  • Type U

Cu-CuNi to DIN 43710

Units

°C or °F

Connection

 
  • Standard connection

1 thermocouple (TC)

  • Generation of average value

2 thermocouples (TC)

  • Generation of difference

2 thermocouples (TC) TC1 – TC2 or TC2 – TC1)

Response time

≤ 250 ms for 1 sensor with open-circuit monitoring

Open-circuit monitoring

can be switched off

Cold junction compensation

 
  • Internal

With integrated Pt100 resistance thermometer

  • External

With external Pt100 IEC 60571 (2-wire or 3-wire connection)

  • External fixed

Cold junction temperature can be set as fixed value

Measuring range

Parameterizable (see table "Digital measuring errors")

Min. measured span

Min. 50 ... 100 °C (90 ... 180 °F) (see table "Digital measuring errors")

Characteristic curve

Temperature-linear or special characteristic

mV sensor

 

Measured variable

DC voltage

Sensor type

DC voltage source (DC voltage source possible over an externally connected resistor)

Units

mV

Response time

≤ 250 ms for 1 sensor with open-circuit monitoring

Open-circuit monitoring

can be switched off

Short-circuit monitoring

can be switched off (value is adjustable)

Measuring range

-10 ... +70 mV
-100   +1100 mV

Min. measured span

2 mV or 20 mV

Overload capability of the input

-1.5 ... +3.5 V DC

Input resistance

≥ 1 MΩ

Characteristic curve

Voltage-linear or special characteristic

Output

 

Output signal

4 ... 20 mA, 2-wire

Auxiliary power supply

11 ... 35 V DC (to 30 V with EEx)

Max. load

(Uaux – 11 V)/0.023 A

Overrange

3.6 ... 23 mA, infinitely adjustable (default range: 3.80 mA ... 20.50 mA)

Error signal (e.g. in the event of sensor breakage)

3.6 ... 23 mA, infinitely adjustable (default value: 22.8 mA)

Sample cycle

0.25 s nominal

Damping

Software filter 1st order 0 ... 30 s (parameterizable)

Protection

Against reversed polarity

Electrically isolated

Input against output (1 kVeff)

Measuring accuracy

 

Digital measuring errors

See Table "Digital measuring errors"

Reference conditions

 
  • Auxiliary power supply

24 V ± 1 %

  • Load

500 Ω

  • Ambient temperature

23 °C

  • Warming-up time

> 5 min

Error in the analog output (digital/analog converter)

< 0.1 % of span

Error due to internal cold junction

< 0.5 °C (0.9 °F)

Temperature effect

< 0.1 % of max. span/10 °C (18 °F)

Auxiliary power supply effect

< 0.005 % of span/V

Effect of load impedance

< 0.012 % of span/100 Ω

Long-term drift

 
  • In the first month

< 0.02 % of max. span

  • After one year

< 0.03 % of max. span

  • After 5 years

< 0.04 % of max. span

Conditions of use

 

Ambient conditions

 

Ambient temperature range

-40 ... +85 °C (-40 ... 185 °F)

Storage temperature range

-40 ... +85 °C (-40 ... 185 °F)

Relative humidity

< 98 %, with condensation

Electromagnetic compatibility

acc. to EN 61326 and NE21

Construction

 

Material

Molded plastic

Weight

50 g (0.11 lb)

Dimensions

See "Dimensional drawings"

Cross-section of cables

Max. 2.5 mm² (AWG 13)

Degree of protection to IEC 60529

 
  • Enclosure

IP40

  • Terminals

IP00

Certificates and approvals

 

Explosion protection ATEX

 

EC type test certificate

PTB 05 ATEX 2040X

  • "Intrinsic safety" type of protection

II 1 G Ex ia IIC T6/T4
II 2(1) G Ex ib/ia IIC T6/T4

  • "Operating equipment that is non-ignitable and has limited energy" type of protection

II 3 G Ex nL IIC T6/T4
II 3 G Ex nA IIC T6/T4

Explosion protection: FM for USA

 
  • FM approval

FM 3024169

  • Degree of protection

IS/Cl I, II, III/Div 1/GP ABCDEFG T6, T5, T4

IS/Cl I/ZN 0/AEx ia IIC T6, T5, T4

NI/Cl I, II, III/Div 2/GP ABCDFG T6, T5, T4

NI/Cl I/ZN 2/IIC T6, T5, T4

Explosion protection to FM for Canada (cFMUS)

 
  • FM approval

FM 3024169C

  • Degree of protection

IS/Cl I/ZN 0/Ex ia IIC T6, T5, T4

NI/Cl I/ZN 2/Ex nAL IIC T6, T5, T4

Software requirements for SIPROM T

 

PC operating system

Windows ME, 2000 and XP; also Windows 95, 98 and 98 SE, but only in connection with RS 232 modem.


Factory setting:

  • Pt100 (IEC 751) with 3-wire circuit
  • Measuring range: 0 ... 100 °C (32 ... 212 °F)
  • Fault current: 22.8 mA
  • Sensor offset: 0 °C (0 °F)
  • Damping 0.0 s

Digital measuring errors

Resistance thermometer

Input

Measuring range

Min. measured span

Digital accuracy

 

°C (°F)

°C

(°F)

°C

(°F)

to IEC 60751

         

Pt25

-200 ... +850 (-328 ... +1562)

10

(18)

0.2

(0.36)

Pt50

-200 ... +850 (-328 ... +1562)

10

(18)

0.15

(0.27)

Pt100 ... Pt200

-200 ... +850 (-328 ... +1562)

10

(18)

0.1

(0.18)

Pt500

-200 ... +850 (-328 ... +1562)

10

(18)

0.15

(0.27)

Pt1000

-200 ... +350 (-328 ... +662)

10

(18)

0.15

(0.27)

to JIS C1604-81

         

Pt25

-200 ... +649 (-328 ... +1200)

10

(18)

0.2

(0.36)

Pt50

-200 ... +649 (-328 ... +1200)

10

(18)

0.15

(0.27)

Pt100 ... Pt200

-200 ... +649 (-328 ... +1200)

10

(18)

0.1

(0.18)

Pt500

-200 ... +649 (-328 ... +1200)

10

(18)

0.15

(0.27)

Pt1000

-200 ... +350 (-328 ... +662)

10

(18)

0.15

(0.27)

Ni 25 to Ni1000

-60 ... +250 (-76 ... +482)

10

(18)

0.1

(0.18)


Resistance-based sensors

Input

Measuring range

Min. measured span

Digital accuracy

 

Ω

Ω

Ω

Resistance

0 ... 390

5

0.05

Resistance

0 ... 2200

25

0,25


Thermocouples

Input

Measuring range

Min. measured span

Digital accuracy

 

°C

(°F)

°C

(°F)

°C

(°F)

Type B

0 ... 1820

(32 ... 3308)

100

(180)

21)

(3.60)1)

Type C (W5)

0 ... 2300

(32 ... 4172)

100

(180)

2

(3.60)

Type D (W3)

0 ... 2300

(32 ... 4172)

100

(180)

12)

(1.80)2)

Type E

-200 ... +1000

(-328 ... +1832)

50

(90)

1

(1.80)

Type J

-210 ... +1200

(-346 ... +2192)

50

(90)

1

(1.80)

Type K

-230 ... +1370

(-382 ... +2498)

50

(90)

1

(1.80)

Type L

-200 ... +900

(-328 ... +1652)

50

(90)

1

(1.80)

Type N

-200 ... +1300

(-328 ... +2372)

50

(90)

1

(1.80)

Type R

-50 ... +1760

(-58 ... +3200)

100

(180)

2

(3.60)

Type S

-50 ... +1760

(-58 ... +3200)

100

(180)

2

(3.60)

Type T

-200 ... +400

(-328 ... +752)

40

(72)

1

(1.80)

Type U

-200 ... +600

(-328 ... +1112)

50

(90)

2

(3.60)


1) The digital accuracy in the range 0 to 300 °C (32 to 572 °F) is 3 °C (5.4 °F).

2) The digital accuracy in the range 1750 to 2300 °C (3182 to 4172 °F) is 2 °C (3.6 °F).

mV sensor

Input

Measuring range

Min. measured span

Digital accuracy

 

mV

mV

μV

mV sensor

-10 ... +70

2

40

mV sensor

-100 ... +1100

20

400


The digital accuracy is the accuracy after the analog/digital conversion including linearization and calculation of the measured value.

An additional error is generated in the output current 4 to 20 mA as a result of the digital/analog conversion of 0.1 % of the set span (digital-analog error).

The total error under reference conditions at the analog output is the sum from the digital error and the digital-analog error (poss. with the addition of cold junction errors in the case of thermocouple measurements).

 

Share |
Video clips
Tin tức mới
Các giấy chứng nhận
Hỗ trợ trực tuyến
Sales Support
0901 185 205
    Support via Yahoo Messanger    
Technical Support
0938 77 40 99
    Support via Yahoo Messanger    
 
Công ty TNHH An Thịnh Tự Động
Trụ sở chính: 186/15 Phan Văn Trị, P.12, Q.Bình Thạnh, TP.HCM
VPGD: 26/1 Nguyễn Bỉnh Khiêm, P. Đa Kao, Q.1, TP HCM
Tel:(84) 28 3910 3379         Fax: (84) 28 3910 3378         Email: info@anthinh.com.vn